Portable Patient Monitor Machine For Continuous Heart Rate Temperature Oxygen Monitoring
Patient Monitor Machine Features
l 12.1 inch color display with 800*600 resolution.
l Three Working Modes: Monitoring, Surgery and Diagnosis.
l Simple and friendly operating display interface.
l Multi-display modes suitable for different applications:
Standard Interface, Large font, ECG Standard Full Display, OXY , Trend Table, BP Trend.
l Built-in rechargeable Lithium battery up to 2 hours of continuous work
l VGA output for external display
Patient Monitor Machine Specification
ECG |
v Lead Mode: 5 Leads (I, II, III, AVR, AVL,AVF, V) v Gain: ´2.5mm/mV, ´5.0mm/mV, ´10mm/mV, ´20mm/mV v Heart Rate: 15-300 BPM (Adult); 15-350 BPM (Neonatal) Resolution: 1 BPM Accuracy: ±1% Sensitivity >200 uV(Peak to peak) v ST measurement range: -2.0 ~ +2.0 mV Accuracy: -0.8mV~+0.8mV: ±0.02mV or ±10%, which is greater Other Range: unspecified v Sweep speed: 12.5 mm/s, 25mm/s, 50mm/s v Bandwidth: n Diagnostic: 0.05~130 Hz n Monitor: 0.5~40 Hz n Surgery: 1~20 Hz |
SPO2 |
v Measuring Range: 0 ~ 100 % v Resolution: 1% v Accuracy: 70% ~ 100% (±2 %) v Pulse Rate: 20-250 BPM v Resolution: 1 BPM v Accuracy: ±3 BPM |
RESP |
v Method: Impedance between RA-LL or RALA v Measurement Range: Adult: 10 ~ 120 BrPM Neonatal / Pediatric: 7 ~ 150 BrPM v Resolution: 1 BrPM v Accuracy: ±2 BrPM |
TEMP |
v Measure and Alarm Range: 0 ~ 50 °C v Resolution: 0.1°C v Accuracy: ±0.1°C |
Standard Parameters |
ECG, RESP, TEMP, NIBP, SPO2, PR |
Optinal Parameters: |
1. Recorder ( Printer) 2. Central monitoring system |
NIBP |
v Method: Oscillometry v Measure mode: Manual, Auto, STAT v Unit: mmHg, kPa v Measure and alarm range: n Adult Mode n SYS 40 ~ 280 mmHg n DIA 10 ~ 225 mmHg n MEAN 20 ~ 240 mmHg n Pediatric Mode n SYS 40 ~ 220 mmHg n DIA 10 ~ 160 mmHg n MEAN 20 ~ 170 mmHg n Neonatal Mode n SYS 40 ~ 135 mmHg n DIA 10 ~ 100 mmHg n MEAN 20 ~ 110 mmHg v Resolution: 1mmHg v Accuracy: ±5mmHg |
RESP | Method: Impedance between RA-LL or RALA Measurement Range: Adult: 10~ 120 BrPM Neonatal /Pediatric:7~150 BrPM Resolution: 1BrPM Accuracy: ±2 BrPM |
Patient Monitor Machine Configuration:
No. |
Item |
Qty |
1 |
Main Unit |
1 |
2 |
5-lead ECG cable |
1 |
3 |
Disposable ECG Electrode |
5 |
4 |
Adult Spo2 probe |
1 |
5 |
Adult NIBP cuff |
1 |
6 |
NIBP extension tube |
1 |
7 |
Temperature probe |
1 |
8 |
Power Cable |
1 |
9 |
User Manual |
1 |
10 |
Fuse |
2 |
11 |
Ground wire |
1 |
Patient Monitor Machine Advantages: